Medical Device Quality Management System (QMS)
This topic is part of the SG Systems Global medical device, ISO 13485, FDA QMSR & lifecycle quality glossary.
Updated December 2025 • ISO 13485, FDA QMSR / 21 CFR 820, EU MDR 2017/745, ISO 14971 • Implants, IVDs, SaMD, capital equipment, disposables & combination products
Medical Device Quality Management System (QMS) is the governance skeleton behind every credible device manufacturer. It’s the integrated set of processes, records, roles and digital systems that prove you can design, manufacture, release, monitor and improve devices without relying on luck and heroics. Done well, a medical device QMS makes design control, risk management, DMR/DHR, UDI, complaints and CAPA feel like one coherent story. Done badly, it’s a pile of procedures nobody follows, CAPAs that never die, and auditors who start asking awkward questions before they’ve even finished the opening meeting.
“If your design file, DMR and DHR tell three different stories about the same device, you don’t have a QMS – you have a paper shield.”
1) What Is a Medical Device QMS?
A Medical Device Quality Management System is the formal framework that describes how a manufacturer ensures devices are consistently safe, effective and compliant over their entire lifecycle. It covers:
- Design & development: design planning, inputs/outputs, verification, validation, usability, software lifecycle (for SaMD) and design transfer.
- Risk management: hazard identification, risk analysis, risk evaluation and risk control per ISO 14971.
- Production & process control: validated processes, equipment, work instructions, in-process checks and environmental controls.
- Records & traceability: DMRs, DHRs, calibration, maintenance, training and batch/serial genealogy.
- Labelling & UDI: correct, approved content, symbols, languages and Unique Device Identification (UDI) assignments and controls.
- Supplier & CMO control: qualification, quality agreements, SCARs and ongoing performance review.
- Post-market surveillance & vigilance: complaints, trend analysis, FSCA/recall, IFU updates and risk/benefit re-evaluation.
Regulators don’t really care what you call it – ISO 13485 system, FDA QMSR, EU MDR QMS – as long as it does these things in a reliable, documented, auditable way and actually shapes reality instead of sitting on a shelf.
2) Regulatory Anchors: ISO 13485, FDA QMSR, EU MDR
Three pillars define what a device QMS must look like globally:
- ISO 13485: International standard for medical device QMS. Focused on risk-based process control, traceability, sterile/clean-critical manufacturing, design control and PMS. It’s what Notified Bodies and many regulators audit against.
- FDA QMSR / 21 CFR 820: The US framework evolving from the Quality System Regulation into a QMSR that more closely aligns with ISO 13485. Key concepts: design controls, production & process controls, CAPA, complaint handling, servicing and DHR/DMR.
- EU MDR / IVDR: Requires a QMS complying with relevant standards, emphasising clinical evaluation, PMS plans, vigilance, UDI and EUDAMED submissions.
A credible Medical Device QMS is effectively your harmonised implementation of all three, with local twists for other markets (Health Canada, UK, Japan, etc.). When an auditor says “show me your QMS”, they’re not asking for a binder; they’re asking how your actual behaviour maps to these rulebooks.
3) Key Elements of a Medical Device QMS
Different standards say it in different ways; the moving parts are remarkably similar:
- Quality policy & objectives: the high-level commitment and measurable goals for quality.
- Organisational structure & responsibilities: management responsibility, independent QA, role definitions.
- Document & record control: SOPs, forms, templates, design files, DMRs, DHRs, electronic records & signatures.
- Design & development controls: planning, inputs, outputs, design reviews, verification, validation, design transfers and design changes.
- Risk management process: integrated ISO 14971 across design, manufacture and PMS, not an isolated FMEA done once for the CER.
- Purchasing & supplier control: qualification, agreements, incoming inspection, SCAR and re-assessment.
- Production & process controls: validated processes, work environments, clean/sterile controls, IPCs and status control.
- Monitoring, measurement, analysis and improvement: audits, complaints, nonconformances, CAPA, trend analysis and management review.
If any of these boxes are “light” – especially design control, DMR/DHR and risk management – you’re running a device business without the core structural supports regulators expect.
4) Design Control, DMR & DHR – The Device QMS Backbone
Three concepts are non-negotiable in a device QMS:
- Design controls: The documented, reviewed, veri-and-validated path from user needs and intended use through design outputs (drawings, SW requirements, circuits, test methods) to a finished, transferable design.
- Device Master Record (DMR): The “how to build it” canonical recipe – BOMs, drawings, specs, manufacturing instructions, inspection procedures and labelling content.
- Device History Record (DHR): The “how we actually built it” batch/serial records – evidence that a specific unit or lot was built according to the DMR and met acceptance criteria.
A Medical Device QMS must make those three things line up. If you can’t show the thread from design file → DMR → DHR for a device – including changes – you’re essentially asking regulators and customers to trust your memory.
5) Risk Management in a Medical Device QMS (ISO 14971)
For devices, risk management isn’t optional fine print; it’s the organising principle. ISO 14971 expects you to:
- identify hazards (mechanical, electrical, software, usability, biological, data, cybersecurity)
- estimate and evaluate risk for each hazard/sequence of events
- implement risk controls (inherent safety, protective measures, information for safety)
- verify and validate risk controls, including potential new risks (risk–risk trade-offs)
- monitor risk in the field and update the risk file over time
A Medical Device QMS has to make this risk file “live” – feeding it with complaints, vigilance reports, usability findings, CAPAs and product changes, not treating it as a one-time input to a CER or 510(k). When auditors ask “show us risk management for this hazard”, they expect to see alignment between risk file, design, DMR, labelling, training and DHRs – not five disconnected stories.
6) UDI, Traceability & Labelling Controls
Modern device QMS implementations must also deal with identity and traceability at a granular level:
- UDI: Assigning and managing device identifiers (DI) and production identifiers (PI) across SKUs, variants, markets and packaging levels.
- Labelling control: Ensuring correct artwork, translations, symbol use, barcode content and versioning for each region and product family.
- Traceability: For implants, sterile products and some IVDs, being able to trace from component lots and processing history to specific devices (and, in some cases, patients), and back.
- Regulatory submissions: Feeding UDI, device and PMS data into regulatory databases (for example EUDAMED) with consistent codes and descriptions.
A Medical Device QMS that treats UDI and labelling as “operations problems” instead of central quality artefacts usually finds out the hard way – via recall, MDR finding or a hospital complaint – that identity mistakes are quality issues, not just logistics annoyances.
7) Complaints, PMS, Vigilance & CAPA
The device QMS doesn’t stop at the loading dock. Core lifecycle elements include:
- Complaint handling: Triaging, investigating and trending complaints, deciding when they are reportable incidents and feeding outcomes back into risk files and design.
- Post-market surveillance (PMS): Proactive collection of performance and safety data – surveys, registry data, field feedback, literature – not just waiting for complaints.
- Vigilance & FSCA: Regulatory reporting of serious incidents and field safety corrective actions; planning and executing removals or updates.
- CAPA: The engine that turns recurring problems into systemic fixes across design, production, suppliers and documentation.
Regulators routinely judge a medical device QMS by its complaint and CAPA files. Superficial investigations, copy-paste root causes and CAPAs that never close are big red flags that the “management” part of the QMS is weak, regardless of how polished the procedures look.
8) What a Medical Device QMS Means for V5
V5 is where the paper QMS and the factory floor can finally speak the same language. Instead of running design controls, production records, CAPA and UDI in separate silos, you can use the V5 stack to make the Medical Device QMS tangible and data-driven.
- V5 Solution Overview
- Provides a unified data model for device SKUs, configurations, components, batches, serials, DMR/DHR objects, suppliers and events.
- Acts as the backbone linking engineering data, shop-floor execution, warehouse status and QMS records into one coherent story per device family.
- V5 MES – Manufacturing Execution System
- Drives execution of DMR content using digital work instructions, routings and recipes that enforce device-specific steps, checks and e-signatures.
- Captures DHR-relevant data (materials, torque values, test results, UDI scans, in-process inspections) at the point of work, creating electronic DHR/eDHR records aligned with ISO 13485 and FDA expectations.
- Enforces risk-based controls from ISO 14971 analyses (for example, dual-verification steps, mandatory tests for certain risk controls, blocked progression when safety-critical steps fail).
- V5 WMS – Warehouse Management System
- Tracks component lots and finished devices by lot/serial, status and location, supporting UDI, implant traceability and “device history” across internal and external movements.
- Implements QC statuses and holds for device components and finished goods, ensuring only released product moves into shipping – and enabling instant identification of stock affected by complaints or FSCA.
- Supports FEFO, shelf-life and environmental constraints for sterile or shelf-limited devices, directly tied into the QMS rules.
- V5 QMS – Quality Management System
- Holds controlled device QMS documentation: design procedures, risk processes, PMS plans, complaint/Vigilance workflows, CAPA and change control, training and competency records.
- Manages device-specific deviations, nonconformances, complaints, FSCA records, SCARs and CAPAs, all linked directly to impacted devices, lots and DHRs.
- Supports ISO 13485 / MDR-style management review through dashboards combining field performance, CAPA status, audit results and production metrics.
- V5 Connect API
- Integrates PLM/ALM (design data), ERP (orders & finance), LIMS (test results), CRM/complaint systems and regulatory submission tools with the V5 execution and quality stack.
- Supports structured exchange with CMOs/contract device manufacturers and key suppliers – for example, sharing DMR extracts, receiving eDHR subsets, or synchronising UDI and labelling data.
In practice, that means your Medical Device QMS is no longer just an ISO 13485 certificate and a set of procedures. It becomes the way V5 runs your device operations: the same logic that auditors read in your QMS manual is visible in your MES flows, WMS holds and QMS records.
9) Implementation Roadmap & Practice Tips
Building or upgrading a Medical Device QMS is heavy work, but you don’t have to do it with a blank sheet. A pragmatic path:
- 1. Map your current QMS reality vs paper. Pick one flagship device or family and trace design → DMR → DHR → complaints → CAPA. Compare what actually happened to what your procedures claim happens.
- 2. Stabilise design control first. Make sure the design file is coherent, risk management is embedded and the DMR is clear enough to be implemented consistently at one site or CMO.
- 3. Clean up DMR/DHR structures. Define standard DMR and DHR templates and minimum content by device class and type. Don’t digitise garbage – fix the structure while you can still see it.
- 4. Implement one eDHR flow in V5. Use V5 MES + V5 QMS + V5 WMS to run one device line fully electronically (or as close as possible), including DHR capture, component traceability and complaint linkage.
- 5. Wire in risk management. Link ISO 14971 risk controls (design and process) to concrete steps in MES, QC plans and labelling; link serious complaints and CAPAs back to the risk file and design history.
- 6. Take UDI & labelling seriously. Standardise how UDI, GTINs and label versions are managed, and use V5 WMS and MES to prevent wrong-label and wrong-UDI events.
- 7. Build usable dashboards. Start with basics: CAPA status by device family, complaint and NC trends, audit findings, eDHR exception rates. Use them in management review, not just as window dressing.
- 8. Tidy the interface with CMOs. Where you outsource device manufacturing, align quality agreements, data exchange and change control with your QMS and V5 data model; stop letting each CMO be its own undocumented universe.
- 9. Iterate device-by-device. Once one reference device flow is solid (design to field), reuse the pattern, templates and V5 configurations for other devices rather than reinventing everything.
The target state is simple to describe: for any device, you can quickly show – with data, not just slides – what you intended to build, how you actually built it, what went wrong, what you did about it and how the field is behaving, with the QMS and V5 telling the same story.
FAQ
Q1. How is a Medical Device QMS different from a general ISO 9001 quality system?
ISO 9001 is a generic quality management standard. A Medical Device QMS under ISO 13485 / FDA QMSR / EU MDR adds device-specific requirements: design controls, DMR/DHR structures, ISO 14971 risk management, sterile/clean environment controls, UDI, PMS, vigilance and often higher expectations for traceability and documentation. A pure ISO 9001 system will not satisfy regulators for medium- or high-risk devices.
Q2. Do all device manufacturers need ISO 13485 certification?
Not always legally, but practically almost always if you want to sell globally or work with reputable partners. Some jurisdictions accept “ISO 13485-equivalent” systems without formal certification, but Notified Bodies, many regulators and larger OEMs treat ISO 13485 as the baseline signal that your QMS is structured properly.
Q3. How does a Medical Device QMS interact with ISO 14971 risk management?
ISO 14971 describes how you manage device risk; the Medical Device QMS describes where and when that process is invoked (design, change control, complaint handling, PMS) and who owns it. A strong QMS ensures risk files are living documents and that risk controls are reflected in DMRs, labelling, procedures, training and DHR evidence – not just stored in a siloed risk management report.
Q4. Can we “bolt on” device QMS requirements to an existing pharma or food QMS?
You can reuse concepts (deviation, CAPA, change control, supplier management), but devices have unique traceability, design control, risk and PMS expectations. In practice, most companies extend an existing QMS with device-specific processes and records, but trying to hide device requirements under a generic QMS without explicit structures for DMR/DHR, UDI and ISO 14971 typically ends badly at audit time.
Q5. How do systems like V5 help with Medical Device QMS compliance?
V5 helps by tying QMS requirements directly into daily operations: DMR content becomes MES recipes and digital work instructions; DHRs are captured automatically as operators scan, measure and sign; quality events and complaints are linked to specific devices and lots; WMS status and traceability rules reflect QMS requirements; and integrations via V5 Connect ensure data doesn’t die in isolated tools. That reduces manual reconciliation, improves data integrity and makes it much easier to demonstrate compliance to auditors and customers.
Related Reading
• Device Foundations: ISO 13485 | ISO 14971 – Medical Device Risk Management | Unique Device Identification (UDI)
• Records & Traceability: Device Master Record (DMR) | Device History Record (DHR) | Lot Traceability & End-to-End Genealogy
• Risk & Events: Quality Risk Management (QRM) | Nonconformance | Deviation | SCAR | CAR | CAPA
• Systems & V5 Platform: Quality Management System (QMS) | V5 Solution Overview | V5 MES – Manufacturing Execution System | V5 QMS – Quality Management System | V5 WMS – Warehouse Management System | V5 Connect API
OUR SOLUTIONS
Three Systems. One Seamless Experience.
Explore how V5 MES, QMS, and WMS work together to digitize production, automate compliance, and track inventory — all without the paperwork.

Manufacturing Execution System (MES)
Control every batch, every step.
Direct every batch, blend, and product with live workflows, spec enforcement, deviation tracking, and batch review—no clipboards needed.
- Faster batch cycles
- Error-proof production
- Full electronic traceability

Quality Management System (QMS)
Enforce quality, not paperwork.
Capture every SOP, check, and audit with real-time compliance, deviation control, CAPA workflows, and digital signatures—no binders needed.
- 100% paperless compliance
- Instant deviation alerts
- Audit-ready, always

Warehouse Management System (WMS)
Inventory you can trust.
Track every bag, batch, and pallet with live inventory, allergen segregation, expiry control, and automated labeling—no spreadsheets.
- Full lot and expiry traceability
- FEFO/FIFO enforced
- Real-time stock accuracy
You're in great company
How can we help you today?
We’re ready when you are.
Choose your path below — whether you're looking for a free trial, a live demo, or a customized setup, our team will guide you through every step.
Let’s get started — fill out the quick form below.